Table 2–10: Warranted characteristics — Signal acquisition system
Name Description
Accuracy, DC Gain TDS 600C: ±1.5% for all sensitivities from 2 mV/div to 10 V/div
± 2.0% at 1 mV/div sensitivity
TDS 500D, 700D, 714L: ±1% for all sensitivities from 1 mV/div to 10 V/div with offset
from 0 V to ±100V
TDS 694C/794D: ±1% for all sensitivities from 10 mV/div to 1 V/div over permitted offset
range
Accuracy, DC Voltage Measurement, Measurement type DC Accuracy
Averaged (using Average mode)
Average of ≥ 16 waveforms
Delta volts between any two averages
of ≥ 16 waveforms acquired under the
same setup and ambient conditions
TDS 600C: ±((1.5% × | reading – Net Offset
1
| )
+ Offset Accuracy) + (0.06 div × V/div))
TDS 500D, TDS 694C, 700D, 714L: ±((1.0% ×
| reading – Net Offset
1
| ) + Offset Accuracy +
(0.06 div x V/div))
TDS 600C: ±((1.5% × | reading | ) + (0.1 div ×
V/div) + 0.3 mV)
TDS 500D, TDS694C, 700D, 714L: ±((1.0% ×
| reading | )
+
(0.1 div x V/div) + 0.3 mV)
Accuracy, Offset
All except TDS 694C/794D
Volts/Div setting
TDS 600C
Offset accuracy
TDS 500D/700D/714L
Offset accuracy
1 mV/div – 100 mV/div ±((0.2% × | Net Off-
set
1
|
)
+ 1.5 mV +
(0.6 div x V/div))
±((0.2% × | Net Off-
set
1
| ) + 1.5 mV +
(0.1 div x V/div))
101 mV/div – 1 V/div ±((0.25% × | Net Off-
set
1
|
)
+ 15 mV +
(0.6 div x V/div))
±((0.25% × | Net Off-
set
1
| ) + 15 mV +
(0.1 div x V/div))
1.01 V/div – 10 V/div
±((0.25% × | Net Off-
set
1
|
)
+ 150 mV +
( 0.6 div x V/div))
±((0.25% × | Net Off-
set
1
| ) + 150 mV +
(0.1 div x V/div))
Accuracy, Offset
TDS 694C and TDS 794D
Volts/Div setting
TDS 694C
Offset accuracy
TDS 794D
Offset accuracy
10 mV/div – 100 mV/div ±((0.2% × | Net Off-
set
1
|
)
+ 1.5 mV +
(0.1 div x V/div))
±((0.2% × | Net Off-
set
1
|
)
+ 1.5 mV +
(0.1 div x V/div))
101 mV/div – 1 V/div ±((2.50% × | Net Off-
set
1
|
)
+ 15 mV +
(0.1 div x V/div))
±((0.25% × | Net Off-
set
1
|
)
+ 15 mV +
(0.1 div x V/div))
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com