Performance Tests
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
141
Turn on the
measurement called
mean and read the
results here.
Figure 16: Measurement of DC accuracy at maximum offset and position
f. Check against limits:
H CHECK that the readout for the measurement Mean readout on
screen is within the limits listed for the current vertical scale and
position/offset/generator settings. Enter value on test record.
H Repeat substep d, reversing the polarity of the position, offset, and
generator settings as is listed in the table.
H CHECK that the Mean measurement readout on screen is within the
limits listed for the current vertical scale setting and position/offset/
generator settings. Enter value on test record.
H Repeat substeps c through f until all vertical scale settings, listed in
Table 13, are checked for the channel under test.
g. Test all channels: Repeat substeps a through f for all four channels.
3. Disconnect the hookup:
a. Set the generator output to 0 V.
b. Disconnect the cable from the generator output at the input connector of
the channel last tested.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Tektronix TDS 784D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 784D and is the answer not in the manual?

Tektronix TDS 784D Specifications

General IconGeneral
BrandTektronix
ModelTDS 784D
CategoryTest Equipment
LanguageEnglish

Related product manuals