Status and Events
TDS200/1000/2000 Series Oscilloscope Programmer Manual
3−1 1
MEASUREMENT:IMMED:VALUE?
Processing Time
Acquiring Waveform Data
ACQUIRE:STATE ON
Figure 3−7: Command processing without using synchronization
The acquisition of the waveform must be completed before the
measurement can be taken on the acquired data. This is achieved by
synchronizing the program so that the measurement command is not
processed by the oscilloscope until the acquisition is complete.
Figure 3-8 shows the desired processing sequence.
MEASUREMENT:IMMED:VALUE?
Processing Time
Acquiring Waveform Data
ACQUIRE:STATE ON
Figure 3−8: Processing sequence with synchronization
Four commands can be used to synchronize the operation of the
oscilloscope with your application program: *WAI, BUSY?, *OPC,
and *OPC?. The *OPC? query is the most simple.
Using the *WAI Command
You can force commands to execute sequentially by using the *WAI
command. This command forces completion of the previous
commands before processing new ones.

Table of Contents

Other manuals for Tektronix TDS2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2002 and is the answer not in the manual?

Tektronix TDS2002 Specifications

General IconGeneral
BrandTektronix
ModelTDS2002
CategoryTest Equipment
LanguageEnglish

Related product manuals