Getting Started
10
TDS1000/2000-Series Digital Oscilloscope User Manual
Self Calibration
The self calibration routine lets you optimize the oscilloscope signal
path for maximum measurement accuracy. You can run the routine at
any time but should always run the routine if the ambient tempera-
ture changes by 5_ Cormore.
To compensate the signal path, disconnect any probes or cables from
the front-panel input connectors. Then, push the UTILITY button,
select the Do Self Cal option and follow the directions on the screen.