Appendix A: Specifications
This appendix contains specifi
cations for the TDS3000C Series
oscilloscopes. All specificatio ns are g uaranteed unless noted as
"typical." Typical specificatio ns are provid e d for your convenience
but are not guaranteed. Speci
fications that are marked with the
symbol are checked in the TDS3000C Specifications and
Performance Verification Technical Reference available on the
www.tektronix.com/manu a
ls Web site.
All specifications apply to all TDS3000C Series models unless noted
otherwise. To meet spec
ifications, two conditions must first be met:
The oscilloscope must have been operating continuously for twenty
minutes within the ope
rating tem perature range specified.
You must perform the Signal Path Compensation operation.
(See page 4, Signal P
ath Compensation (SPC).) If the operating
temperature changes by mo re than 1 0 °C (18 °F), y ou m ust perform
the Signal Path Compensatio n opera tion again.
Acquisition
Acquisition
modes
Sample (Normal), Pe ak detect, Enve lope, and Average
Acquisition mode
Acquisition stops after
Sample, Peak De tect One acquisition, all
channels simultaneously
Single
Sequence
Average, Envelope N acquisitions, all channels
simultaneously, N is
settable from 2 to 256
(or ∞ for Envelope)
TDS3000C Series O scilloscope User Manual 149