Appendix A: Specifications
Inputs
Measured on one channel, with test signal applied to another
channel, and with the same scale and coupling settings on
each chann el
Frequency range
Crosstalk
≤ 100 MHz ≥ 100:1
≤ 300 MHz ≥ 50:1
Channel
to channel
crosstalk,
typical
≤ 500 MHz ≥ 30:1
Differential
delay,
typical
100 ps between any two channels with the same scale and
coupling settings
1
For definitions, refer to C ompliance Information.(Seepagexi,Compliance information.)
TDS3000C Series O scilloscope User Manual 151