PS2520 Series Performance Verification
Handheld and Benchtop Instruments Basic Service
11
20. Press SHIFT
STORE 1 to store the above values in the memory 01
location. Verify that the MEMORY, CURRENT (A), and VOLTS readouts
display the values listed in step 19 for memory location 01.
21. Reset the power supply as follows:
VOLTS SET 6 V
CURRENT SET 3 A
DELAY 3 s
22. Press SHIFT
STORE 2 to store the above values in the memory 02
location. Verify that the MEMORY, CURRENT (A), and VOLTS readouts
display the values set in step 21 for memory location 02.
23. Repeat steps 6 through 12 above.
24. Press STEP SET. Set the VOLTS SET to 1 V.
25. Press VOLTS
and verify that as you attempt to decrease the voltage
below 0.00 V, “Err - 018” appears on the CURRENT (A) and VOLTS
readouts. The voltage should decrease in 1 volt steps on the VOLTS readout.
26. Press VOLTS
and verify that as you attempt to increase the voltage past
6.50 V, “Err - 016” appears on readouts. The voltage should increase in
1 volt steps on the VOLTS readout.
27. Press STEP SET. Set the CURRENT SET to .1 A.
28. Press CURRENT
and verify that as you attempt to decrease the current
below 0.000 A, “Err - 019” appears on the readouts and the “C.C.” indicator
lights. The current should decrease in .1 ampere steps on the CURRENT (A)
readout.
29. Press CURRENT
and verify that as you attempt to increase the current
past 3.100 A, “Err - 017” appears on the readouts. The current should
increase in .1 ampere steps on the CURRENT (A) readout.
OCP, Series, and Parallel Mode Indicators. Use the following steps to verify OCP,
series, and parallel output indicator function.
1. On the power supply front panel, press OCP ON/OFF to enable the OCP.
Verify that the “OCP” indicator lights up on the display.
2. Press OCP ON/OFF. Verify that the “OCP” indicator turns off.
3. Press SHIFT
SERIES/INDEP to configure the power supply outputs for
series operation. Verify that the “SERIES” indicator lights up on the display.
4. Press SHIFT
PARA/INDEP to configure the power supply outputs for
parallel operation. Verify that the “PARA” indicator lights up on the display.

Table of Contents

Other manuals for Tektronix THM565

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix THM565 and is the answer not in the manual?

Tektronix THM565 Specifications

General IconGeneral
BrandTektronix
ModelTHM565
CategoryMultimeter
LanguageEnglish

Related product manuals