EasyManuals Logo
Home>TSI Instruments>Laboratory Equipment>3082

TSI Instruments 3082 User Manual

TSI Instruments 3082
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #130 background imageLoading...
Page #130 background image
5-42 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
8. When all the data has been added, again press the currently selected
setpoint button to stop flow.
9. The Calibrate button becomes active. To recalibrate the flow rate,
press Calibrate. The Classifier column is recalculated based on the
new calibration curve. The Error (%) column indicates the accuracy of
the new calibration curve. An onscreen message tells you the
calibration is complete and has been saved to the instrument.
Figure 5-27
Electrostatic Classifier User Calibration ScreenCalibration Saved to
Instrument
10. Touch OK to save or Clear to the data.
Check Existing Calibration
If you do not want to re-calibrate flow, but instead want to check the
accuracy of your current calibration, use the Check button to verify an
existing calibration. To check a calibration, follow these instructions:
1. For best results (more stable flow), install a DMA to add back pressure
to the system.

Table of Contents

Other manuals for TSI Instruments 3082

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments 3082 and is the answer not in the manual?

TSI Instruments 3082 Specifications

General IconGeneral
BrandTSI Instruments
Model3082
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals