6-6 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
3. Using a small slotted screwdriver, remove the O-ring.
C a u t i o n
Do not puncture the O-ring.
4. Grease the replacement O-ring 6006389 and insert into the groove in
the classifier inlet.
Greasing O-rings
Greasing the radial-seal O-rings extends their life. The impactor O-rings
should be greased once a year as part of your routine maintenance
procedures.
To grease the O-rings, follow these instructions:
1. Remove the O-rings from the impactor following the procedure noted
above.
2. Place a small amount of standard vacuum grease on your finger.