‘Calibration Mode’ Mode will shown by pressing key, after Mode display.Setp
4.3 Calibration Mode
Range / Values
Parameter
CAL
0000
CALIBRATION MODE
Password for Calibration Mode Parameters
Display
Scroll the digit by key, Increase by key.
Range : Values between to0000 9999
Default Value : 0000
CALz
----
CALIBRATION ZERO or LOW INPUT
Calibration Zero / Low at Low / Zero PV Input.
NOTE : After Zero Cal. press key for next Para.
Apply Low signal [as per input] & press key.
Range :
Default Value : ----
According to Input Type
CALS
----
CALIBRATION SPAN or HIGH INPUT
Calibration Span / High at Span / High PV Input.
NOTE : After Span Cal. press key for next para.
Apply High signal [as per input] & press key.
CALL
strt
CALIBRATION LOW [RETRANSMISSION] OUTPUT
Calibration desired Low Output at PV Display.
NOTE : After out low cal. press key for next Para.
Set output by , & key, press key
for desired low measured output achieved.
Note : After Calibration random value
appeared in display, as per input.
Refer below Table for Cal. values range.
Note : After Calibration random value
appeared in display, as per input.
Refer below Table for Cal. values range.
Range :
Default Value : ----
According to Input Type
Note : After Cali. random value should be
below appeared in display.1000
Range :
Default Value : -> to start cal.strt
According to Output Type
CALH
strt
CALIBRATION HIGH [RETRANSMISSION] OUTPUT
Calibration desired High Output at PV Display.
NOTE : After out low cal. press key for next Para.
Set output by , & key, press key
for desired high measured output achieved.
Range :
Default Value : -> to start cal.strt
According to Output Type
Note : After Cali. random value should be
below appeared in display.8000
CALa
0032
CALIBRATION AMBIENT [THERMO COUPLE INP.]
Ambient adjustment for compensation [T/C Inp.]
NOTE : AfterAmb.Cal.press keyforexit.
Set Ambient Temp by & key,press key.
Note : After Calibration random value
appeared in display, as per input.
Range :
Default Value : 0032
As per Ambient Temperature.
0100 0600~
0200 1000~
0200 1000~
0200 1000~
0200 1000~
0200 1000~
7000 8000~
7000 8000~
7000 8000~
7000 8000~
7000 8000~
7000 8000~
INPUT
J-T/C
K-T/C
R-T/C
S-T/C
T-T/C
E-T/C
0mV
0mV
0mV
ZERO VALUE
FEED VALUE COUNT
SPAN VALUE
FEED VALUE COUNT
0mV
0mV
0mV
18 mV
18 mV
42 mV
42 mV
42 mV
18 mV
INPUT
ZERO VALUE
FEED VALUE COUNT
SPAN VALUE
FEED VALUE COUNT
0100 0600~
0100 1000~
0100 1000~
0100 1000~
0900 1500~
7000 8000~
7000 8000~
7000 8000~
7000 8000~
2000 3000~
4 ~ 20mA
0 ~ 20mA
1~5VDC
10VDC
PT-100
4mA
0mA
1VDC
0VDC
20 mA
20 mA
5VDC
10 VDC
CALIBRATION VALUES :
0°C
[100.0Ω]
100°C
[138.51Ω]