5. Operation
5.1 Probe Start-Up
When powered on, the probe starts up within 2 seconds and the digital/analog outputs are
activated. The probe performs a start-up purge, which takes approximately 4 minutes. During
the purge, hydrogen peroxide, relative humidity, and relative saturation measurements are not
available. If the probe is in analog mode, analog outputs are in the error state (default: 3.6 mA)
during the purge.
Measurements from the outputs (digital and analog) will reach
specified accuracy after a 8½-
minute warm-up period. For this reason, you should design your system so that it does not rely
on measurements from the probe during this time.
More Information
Behavior at Exposure to H2O2 (page 25)
5.2
 Behavior at Exposure to H
2
O
2
When there is H
2
O
2
in the probe's environment, the probe must
always be powered on. When powered on, the PEROXCAP sensor is heated,
which permits using the probe in condensing H
2
O
2
conditions, maintains
measurement performance, and lengthens the probe's lifetime. When the probe
is powered o, exposure to H
2
O
2
condensation can break the PEROXCAP sensor
within a day, and the sensor will not recover.
CAUTION!
When the bio-decontamination process starts and the probe is exposed to H
2
O
2
, the probe's
H
2
O
2
concentration reading changes to > 0 ppm after approximately 20 ... 30 seconds. This
time is included in the response time of the probe.
5.3
 H
2
O
2
Concentration Reading When Not
Exposed to H
2
O
2
The PEROXCAP sensor consists of two humidity sensors that have a minor dierence in
behavior when the humidity level changes. Because of this
dierence, the H
2
O
2
concentration
reading may vary slightly (typically 0 … 3 ppm) when the probe is not exposed to H
2
O
2
. This
variation is normal and does not require any actions. If needed, you can hide the variation by
setting a low clipping limit with the Vaisala Insight software, Modbus
configuration register
030C
hex
, or an Indigo 200 transmitter. For example, if you set the low clipping limit to 3 ppm,
any reading < 3 ppm is clipped to show 0 ppm.
Chapter 5 – Operation
25