Register Description Data
Format
Unit
260 0103
hex
Relative humidity 16-bit
signed
integer
%RH *100
261 0104
hex
Absolute hydrogen peroxide 16-bit
signed
integer
mg/m
3
264 0107
hex
Water concentration by volume 16-bit
signed
integer
ppm
265 0108
hex
Water vapor pressure 16-bit
signed
integer
hPa
266 0109
hex
Absolute humidity (water) 16-bit
signed
integer
g/m
3
267 010A
hex
Water vapor saturation pressure 16-bit
signed
integer
hPa
268 010B
hex
Temperature from the temperature
probe
16-bit
signed
integer
°C * 100
A.4.2 Configuration Registers
Default power-up values (registers 773 ... 775) are written into non-
volatile EEPROM memory. The EEPROM memory implementation has a limit of
30000 writes, and is intended to be used only when saving long-term or
permanent configurations. Use the volatile memory (registers 769 ... 771, values
cleared on power-up) for non-permanent configurations (for example, if the
probe is used in a system that regularly updates the given values).
CAUTION!
HPP272 User Guide M211972EN-B
48

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala HPP272 and is the answer not in the manual?

Vaisala HPP272 Specifications

General IconGeneral
BrandVaisala
ModelHPP272
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals