11. Cleaning and Disinfection
126 vatech A9 (Model name: PHT-30CSS) User Manual
Do not use spray cleaners or solvents as they could flow into the equipment and
damage the electrical components or cause a fire.
Do not use any cleaning products which contain silicon. They could potentially
damage the equipment’s electrical components.
The following table summarizes the standard cleaning procedures to be performed by
the operator.
Components Cleaning Process
Bite
(Normal Bite, Special Bite
A, and Special Bite B)
Clean with non-alcohol-based, non-corrosive
cleaning solution and gently wipe with a dry towel
before the next patient.
Chinrest
Clean with non-alcohol-based, non-corrosive
cleaning solution and gently wipe with a dry towel
before the next patient.
Computer and peripherals
Follow the manufacturers’ instructions found in the
accompanying manuals.
Outer covers of equipment Wipe the unit with a dry cloth at the end of each day.
Do not use cleaning agents in aerosol or spray form directly
on the surface of the equipment.
11.2 Disinfection
Use only disinfectants that comply with the valid requirements of the respective
national regulatory body or whose bactericidal, fungicidal, and virucidal properties
have been verifiably tested and approved accordingly.
Sterilization and disinfection should be performed thoroughly for items that have
been in frequent contact with patients and operators.
Do not use UV systems to disinfect the equipment, as exposed parts of the
equipment can turn yellow or discolor.
The use of unsuitable Cleaners or disinfectant agents and methods can damage
the equipment and accessories. Only use the Cleaners or disinfectant agents
specified or approved by VATECH
The following Cleaners or disinfectant agents have been evaluated for safe use
on the surfaces.
Never combine products or liquids other than the products listed above.
Damages to surfaces and materials due to the use of different products cannot be
excluded even if they are not included in the exceptions mentioned above.
Use a non-alcoholic chlorine dioxide-based disinfectant.

Table of Contents

Other manuals for Vatech A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech A9 and is the answer not in the manual?

Vatech A9 Specifications

General IconGeneral
BrandVatech
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals