PaX-Primo
10
1.5. PaX-Primo’s general specification
z X-ray beam: Fan Beam
z Reconstruction Algorithm: A real time reconstructing algorithm
z Dynamic Range: 14 bit
z Scan Time (sec) at normal: 9.7 sec (typical).(2~13.5 sec.)
z Rotating Unit Scan Angle (degree): 240
Actual rotating angle around the object
z X-ray Exposure Angle (degree): 220
Actual spanning angle to expose the object to X-Ray: In fact the instant that the rotating unit
turns by 10 degrees, the X-Ray exposure starts and stops at 10 degrees before the rotating unit
reaches the full rotation: 240-10-10=220
z Patient Position: Stand
z Patient Alignment: 3 guiding light beams
z Reconstruction Time: About 45sec
z Focal spot size: 0.35 X0.5 mm
z FDD (SID): 531.2 mm
z Magnification: 1.31
z Sensor type: CMOS (Area mode)
z Pixel pitch: 100um
z Image pixel size: 100um
z Active pixel array: 1504x60
z Active area: 150.4x6.0mm
z Data acquisition speed: 1 Gb
z Weight: 75 kg
Exposure angle
Rotating unit spanning angle

Table of Contents

Other manuals for Vatech PaX-Primo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PaX-Primo and is the answer not in the manual?

Vatech PaX-Primo Specifications

General IconGeneral
BrandVatech
ModelPaX-Primo
CategoryMedical Equipment
LanguageEnglish

Related product manuals