Installation manual
3
Guide Lines for Protection against Radiation
The X-ray system may cause injury to the patients if used improperly. The instructions contained in this
manual must be read and followed when operating PaX-Primo. The world standard regulations pertaining
to radiation safety must be observed.
When exposing X-ray, User must be behind the protective wall, or take other protective actions. When a
breakdowns or troubles appear, User keeps at least 2m (7feet) away from the X-ray system to release the
exposure switch while observing patient and capture-progress.
User must provide the protective clothes to the patient. Before capturing, pregnant women must always
consult with doctors.
X-ray Room Space for System Installation
Recommended Minimum Space
z PaX-Primo 2000(L) * 2000(W) * 2500(H) mm
- Above space is considered for the movement of both system operator and patient.
- The system is normally installed beside a wall, and operator uses the system on left.
Width of Door
- The width of door is more than 800mm for system movement into X-ray room.
Ground
- The condition of ground should be flat for system balance.
- Ground should support a min. 500KG of weight.
Power Supply
- For system stability please guarantee 3KW power supply.
Protection against radiation
- For protection against radiation follow the government or local standards.
Safety Zone
- Check the safety zone.
Do not use this system in any place where there is an explosive danger.