27
4. OPERATION PROCEDURE
Programmed Operation
PROGRAM mode runs a combination of times and temperatures in a series of programmed
steps as one operation. See below.
Running programs
1
Turn power ON
Turn main power switch ON (
) (idle).
Press and hold to turn control panel power on
(standby).
Current temperature reading is shown in upper display,
Temperature setting is shown in lower display.
Fan begins running (Runs while door is closed and
stops whenever door is opened).
! Programs must be entered prior to starting a programmed operation run.
For details on entering programs, see “Programming Procedure” on P.30
Create as many as 99 programs and steps in total (i.e. 11 programs with a maximum of 9 steps
each). Step repeats are not counted in this total.
2
Select program mode
Press repeatedly until PROGRAM lamp lights.
Lower display shows XX (“XX” signifies program
number last used. Factory default is “ ).
Constant Temperature (FIXED TEMP) is factory
default mode. Once mode has been changed,
the last mode run will be selected on
subsequent startups.