Your Accu-Chek Spirit Combo insulin pump
43
Hold your Pump diagonally with the adapter
pointing down. Insert the cartridge into the car-
tridge compartment.
The cartridge is correctly inserted when the end
plate of the piston rod is correctly seated against
the cartridge plunger.
Twist (do not push) the adapter clockwise until it
is correctly seated against the cartridge compart-
ment.
If the end plate of the piston rod is not correctly
aligned with the cartridge plunger, remove the
cartridge from the cartridge compartment and
repeat steps 1. to 6.