102 14 Ventilator service and maintenance intervals
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
Every 4 years:
Replace the following components:
Air and Oxygen Input Filters
Battery Pack
Calibration Valves
Cooling Air Filter (casing bottom)
Exhalation membrane
Flow sensor
Input pressure regulators pressurized Air / Oxygen
Proportional Valves pressurized Air / Oxygen mixer
Lithium Battery (real time clock battery)
O
2
sensor
14.1 Disposal
fabian must not be disposed of with household waste but must be handed in
for disposal as electrical and electronic equipment, including battery /
batteries according to applicable national, regional and local rules, laws and
regulations.
Accessories and consumables must be disposed of in accordance with the
relevant instructions for use.
For information, contact your local environmental or regulatory agency, or
an appropriate waste disposal company.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian +nCPAP evolution and is the answer not in the manual?

ACUTRONIC fabian +nCPAP evolution Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian +nCPAP evolution
CategoryMedical Equipment
LanguageEnglish

Related product manuals