18 Special procedures 120
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
18 Special procedures
18.1 Connect Nitric Oxide (NO) tubing sets
IMPORTANT:
Only INOmax and INOvent NO systems are approved by ACUTRONIC Medical Systems.
If other manufacturers’ NO systems are used, they must be validated first before use with
the fabian.
If Nitric Oxide Systems like the INOmax or INOvent are used to deliver NO, make sure the setup
of the circuit is done as per below graphic. It is important to place the flow sensor of the NO
delivery system directly at the fresh gas outlet of the fabian +nCPAP evolution.
The measuring line must be placed close to patient in Inspiratory Limb.
fabian Ventilator
Patient Gas Sample Line with Nafion
INOmax NO System (alternatively,
INOvent can be used, too)
NO/N
2
Injector Tube
Connecting Tube (15 inches)
15M x 15M Adapter
Injector Module Electrical Cable
Injector module
22F x 15M Adapter
Humidifier Inlet
Humidifier
Humidifier Outlet
Inspiratory Breathing Circuit Hose
Gas Sample Tee
Patient Wye
Proximal Pressure Tube
Expiratory Breathing Circuit Hose

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian +nCPAP evolution and is the answer not in the manual?

ACUTRONIC fabian +nCPAP evolution Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian +nCPAP evolution
CategoryMedical Equipment
LanguageEnglish

Related product manuals