8 Battery operation 71
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
8.1 Power failure
In the event of a Power Failure, the device
will automatically switch to Battery power
without interruption.
A visual and audible message will be
triggered.
Power failure
This message can be confirmed
with the Alarm Silence button.
The device will continuously monitor the battery run time.
Additional messages will be triggered at a remaining Run Time of:
15 minutes
30 minutes
60 minutes, only if the remaining run time is less than 60 minutes during a Power
Failure
If the remaining Run Time is less than one
minute, a warning will appear prompting to
immediately switch to a Primary Power
Source.
After 60 minutes on Battery power, the
Battery is drained, and fabian will power
OFF.
Depleted Battery
Connect to Main
Charge battery for approx. 5 hours