EasyManuals Logo
Home>ACUTRONIC>Medical Equipment>fabian +nCPAP evolution

ACUTRONIC fabian +nCPAP evolution User Manual

ACUTRONIC fabian +nCPAP evolution
149 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
46 6 Configurations menu
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
The Flow Sensor Calibration needs to be performed each time:
A new sensor is put in place.
After device startup
After enabling the flow sensor (manual and automatic)
After patient range change (note the sensor type must match the
patient range selected)
After reconnection of a flow sensor resolving a disconnection alarm
We recommend cleaning the Flow Sensor once daily.
If Zero Flow and Tidal Volume measurements are detected in numeric and
graphs after:
Disconnection
High Leaks
Low-level Rain-out
Over-breathing
flow measurement will be automatically re-started after 6 seconds.
If Flow and Tidal Volume measurements are not regained, consider the
following:
Checking ET tube position (high leaks at the patient) and blockage
(during or after in-line suction, or after surfactant therapy)
Flow Sensor Leakage (loose flow insert or leak)
Flow Sensor Contamination (single-use sensor recommended)
Rain-out (contact ACUTRONIC Medical Systems sales for
recommended Patient Circuits and Humidifier to prevent Rain-out)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian +nCPAP evolution and is the answer not in the manual?

ACUTRONIC fabian +nCPAP evolution Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian +nCPAP evolution
CategoryMedical Equipment
LanguageEnglish

Related product manuals