Service
Analog Bandwidth - Maximum Frequency Check
4-13
Analog Bandwidth - Maximum Frequency Check
CAUTION Ensure that the input voltage to the oscilloscope never exceeds 300 Vrms.
Specification
Equipment Required
Analog Bandwidth (-3 dB)
DSO3062A 60 MHz
DSO3102A 100 MHz
DSO3152A 150 MHz
DSO3202A 200 MHz
Description Critical Specifications Recommended Model/
Part Numbers
Signal Generator 100 kHz to 1 GHz at 200 mVrms Agilent 8648A
Power Splitter outputs differ by < 0.15 dB Agilent 11667B
Power Meter Agilent E-series with power sensor compatibility Agilent E4418B
Power Sensor 100 kHz to 1 GHz ±3% accuracy Agilent 8482A
SMA Cable SMA (m) to SMA (m) 24 inch
Adapter 50
BNC feed through terminator
Adapter Type N (m) to SMA (f) Agilent 1250-1250
Adapter Type SMA (m) to BNC (m) Agilent 1250-0831