<function>:RESolution {<resolution>|MINimum|MAXimum}
Select the resolution for the specified function (not valid for frequency,
period, or ratio). Specify the resolution in the same units as the
measurement function, not in number of digits.
MIN selects the smallest
value accepted, which gives the most resolution.
MAX selects the largest
value accepted which gives the least resolution. [Stored in volatile memory]
<
function>:RESolution? [MINimum|MAXimum]
Query the resolution for the selected function. For frequency or period
measurements, the multimeter returns a resolution setting based upon
a 3 Hz input frequency.
<
function>:NPLCycles {0.02|0.2|1|10|100|MINimum|MAXimum}
Select the integration time in number of power line cycles for the
present function (the default is 10
PLC). This command is valid only for
dc volts, ratio, dc current, 2-wire ohms, and 4-wire ohms.
MIN = 0.02.
MAX = 100. [Stored in volatile memory]
<
function>:NPLCycles? [MINimum|MAXimum]
Query the integration time for the selected function.
FREQuency:APERture {0.01|0.1|1|MINimum|MAXimum}
Select the aperture time (or gate time) for frequency measurements
(the default is 0.1 seconds). Specify 10 ms (4
1
2
digits), 100 ms (default;
5
1
2
digits), or 1 second (6
1
2
digits). MIN = 0.01 seconds. MAX = 1 second.
[Stored in volatile memory]
FREQuency:APERture? [MINimum|MAXimum]
Query the aperture time for frequency measurements.
PERiod:APERture {0.01|0.1|1|MINimum|MAXimum}
Select the aperture time (or gate time) for period measurements
(the default is 0.1 seconds). Specify 10 ms (4
1
2
digits), 100 ms (default;
5
1
2
digits), or 1 second (6
1
2
digits). MIN = 0.01 seconds. MAX = 1 second.
[Stored in volatile memory]
PERiod:APERture? [MINimum|MAXimum]
Query the aperture time for period measurements.
Chapter 4 Remote Interface Reference
Measurement Configuration Commands
122

Table of Contents

Other manuals for Agilent Technologies 34401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34401A and is the answer not in the manual?

Agilent Technologies 34401A Specifications

General IconGeneral
BrandAgilent Technologies
Model34401A
CategoryMultimeter
LanguageEnglish

Related product manuals