[SENSe:]DETector:BANDwidth {3|20|200|MINimum|MAXimum}
Specify the lowest frequency expected in the input signal. The multimeter
selects the slow, medium (default), or fast ac filter based on the frequency
you specify.
MIN = 3 Hz. MAX = 200 Hz. [Stored in volatile memory]
[SENSe:]DETector:BANDwidth? [MINimum|MAXimum]
Query the ac filter. Returns “+3.000000E+00”,+2.000000E+01”, or
+2.000000E+02”.
[SENSe:]ZERO:AUTO {OFF|ONCE|ON}
Disable or enable (default) the autozero mode. The
OFF and ONCE
parameters have a similar effect. Autozero OFF does not issue a new
zero measurement until the next time the multimeter goes to the
wait-for-trigger” state. Autozero
ONCE issues an immediate zero
measurement. [Stored in volatile memory]
[SENSe:]ZERO:AUTO?
Query the autozero mode. Returns 0” (
OFF or ONCE) or “1” (ON).
INPut:IMPedance:AUTO {OFF|ON}
Disable or enable the automatic input resistance mode for dc voltage
measurements. With
AUTO OFF (default), the input resistance is fixed
at 10 M
for all ranges. With AUTO ON, the input resistance is set to
>10 G
for the 100 mV, 1 V, and 10 V ranges. [Stored in volatile memory]
INPut:IMPedance:AUTO?
Query the input resistance mode. Returns “0” (
OFF) or “1” (ON).
ROUTe:TERMinals?
Query the multimeter to determine if the front or rear input terminals
are selected. Returns “
FRON or REAR”.
4
Chapter 4 Remote Interface Reference
Measurement Configuration Commands
123

Table of Contents

Other manuals for Agilent Technologies 34401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34401A and is the answer not in the manual?

Agilent Technologies 34401A Specifications

General IconGeneral
BrandAgilent Technologies
Model34401A
CategoryMultimeter
LanguageEnglish

Related product manuals