EasyManuals Logo
Home>Agilent Technologies>Test Equipment>8647A

Agilent Technologies 8647A User Manual

Agilent Technologies 8647A
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #380 background imageLoading...
Page #380 background image
8-62
Performance Tests
8648A Test Record
Table 8-11. Phase Modulation Distortion Performance Test
Frequency
(MHz)
Deviation
(Rad)
Limits (%)
Uncertainty
(%)Lower Measured Upper
0.250 5 0 1 0.26
100 5 0 1 0.26
400 5 0 1 0.26
500 5 0 1 0.26
750 5 0 1 0.26
1000 5 0 1 0.26
Table 8-12. Residual FM Performance Test
RF Frequency
8648A
(MHz)
LO Frequency
8663A
(MHz)
Input Frequency
8902A
(MHz)
Limits (Hz)
Uncertainty
(±Hz)Upper Measured
10 11.5 700 7 0.45
100 101.5 700 7 0.45
248 249.5 700 7 0.45
251 252.5 700 4 0.35
400 401.5 700 4 0.35
501 502.5 700 7 0.45
750 751.5 700 7 0.45
1000 1001.5 700 7 0.45