1a-28
Operation Examples
Signaling a Numeric-Type FLEX Pager (Option 1EP Only)
9. ADDRESS TYPE and ADDRESS1 are set automatically when the capcode is entered in the
previous menu. If A0012477 was entered, SHORT and 0045245 would be displayed
respectively. Generally, you would not change these settings.
Press the (NEXT) key to move to the next page.
Selecting the Protocol Settings
10.FRAME, PHASE, and COLLAPSE CYCLE are also set automatically when you enter the
pager’s capcode. If you entered A0012477 previously, 011, D, and 4 would be displayed.
NOTE If MODE is set to BURST or CONT instead of SINGLE, you may choose to change
the collapse cycle to 1 to 7. With a collapse cycle of 4, the pager will receive
the message once every 16 frames (2
4
). If the collapse is 0 or 1, the pager will
receive the message either every frame (2
0
), or every other frame (2
1
).
Press the (NEXT) key to move to the next page.

Table of Contents

Other manuals for Agilent Technologies 8647A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8647A and is the answer not in the manual?

Agilent Technologies 8647A Specifications

General IconGeneral
BrandAgilent Technologies
Model8647A
CategoryTest Equipment
LanguageEnglish

Related product manuals