EasyManuals Logo
Home>Agilent Technologies>Test Equipment>8647A

Agilent Technologies 8647A User Manual

Agilent Technologies 8647A
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #456 background imageLoading...
Page #456 background image
8-138
Performance Tests
8648D Test Record
NOTE If the automated Power Level Accuracy performance test is performed, the
RF Level Accuracy performance test is not required.
Table 8-87. RF Level Accuracy Performance Test - Part 1
Frequency
(MHz)
Amplitude
(dBm)
Limits (dBm)
Uncertainty
(±dB)Lower Measured Upper
31312 140.22
10 9.0 11.0 0.22
4 3.0 5.0 0.22
5.9 6.9 4.9 0.22
15.9 16.9 14.9 0.23
25.9 26.9 24.9 0.23
35.9 36.9 34.9 0.23
45.9 46.9 44.9 0.23
55.9 56.9 54.9 0.24
65.9 66.9 64.9 0.24
75.9 76.9 74.9 0.25
85.9 86.9 84.9 0.25
95.9 96.9 94.9 0.25
96.0 97.0 95.0 0.25
105.9 106.9 104.9 0.26
106.0 107.0 105.0 0.26
115.9 116.9 −114.9 0.27
116.0 117.0 115.0 0.27
125.9 126.9 124.9 0.28
126.0 127.0 125.0 0.29
127.0 128.0 126.0 0.29

Table of Contents

Other manuals for Agilent Technologies 8647A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8647A and is the answer not in the manual?

Agilent Technologies 8647A Specifications

General IconGeneral
BrandAgilent Technologies
Model8647A
CategoryTest Equipment
LanguageEnglish

Related product manuals