EasyManuals Logo
Home>Agilent Technologies>Inverter>E8257D

Agilent Technologies E8257D User Manual

Agilent Technologies E8257D
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #183 background imageLoading...
Page #183 background image
Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide 173
Programming the Status Register System
Status Groups
Data Questionable Frequency Event Enable Register
Lets you choose which bits in the Data Questionable Frequency Event Register set the summary bit
(bit 5 of the Data Questionable Condition Register) to 1.
Command:
STATus:QUEStionable:FREQuency:ENABle <value>, where <value> is the sum of the decimal values
of the bits you want to enable.
Example: Enable bit 4 and bit 3 so that whenever either of those bits are set to 1, the Data Questionable
Frequency summary bit of the Data Questionable Condition Register is set to 1. Send the command
STAT:QUES:FREQ:ENAB 520 (16 + 8).
Query: STATus:QUEStionable:FREQuency:ENABle?
Response: Decimal value of the sum of the bits previously enabled with the
STATus:QUEStionable:FREQuency:ENABle <value> command.

Table of Contents

Other manuals for Agilent Technologies E8257D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8257D and is the answer not in the manual?

Agilent Technologies E8257D Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8257D
CategoryInverter
LanguageEnglish

Related product manuals