Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide 81
Programming Examples
GPIB Programming Interface Examples
printf("Example program to set up the signal generator\n");
printf("for an AC-coupled FM signal\n");
printf("Press any key to continue\n");
printf("\n");
getch();
printf("\n");
viPrintf(vi, "*RST\n"); // Resets the signal generator
viPrintf(vi, "FM:SOUR EXT2\n"); // Sets EXT 2 source for FM
viPrintf(vi, "FM:EXT2:COUP AC\n"); // Sets FM path 2 coupling to AC
viPrintf(vi, "FM:DEV 20 kHz\n"); // Sets FM path 2 deviation to 20 kHz
viPrintf(vi, "FREQ 700 MHz\n"); // Sets carrier frequency to 700 MHz
viPrintf(vi, "POW:AMPL -2.5 dBm\n"); // Sets the power level to -2.5 dBm
viPrintf(vi, "FM:STAT ON\n"); // Turns on frequency modulation
viPrintf(vi, "OUTP:STAT ON\n"); // Turns on RF output
// Print user information
printf("Power level : -2.5 dBm\n");
printf("FM state : on\n");
printf("RF output : on\n");
printf("Carrier Frequency : 700 MHZ\n");
printf("Deviation : 20 kHZ\n");
printf("EXT2 and AC coupling are selected\n");
printf("\n"); // Prints a carrage return
// Close the sessions
viClose(vi);
viClose(defaultRM);
}
Generating an Internal FM Signal Using VISA and C
In this example the VISA library is used to generate an internal FM signal at a carrier frequency of
900 MHz and a power level of −15 dBm. The FM rate will be 5 kHz and the peak deviation will be
100 kHz. Launch Microsoft Visual C++ 6.0, add the required files, and enter the following code into
your .cpp source file. visaex6.cpp performs the following functions:
•error checking
• resets the signal generator
• sets up the signal generator for FM path 2 and internal FM rate of 5 kHz
• sets up FM path 2 deviation to 100 kHz
• sets carrier frequency to 900 MHz
• sets the power level to - 15 dBm
• turns on frequency modulation and RF output
The following program example is available on the signal generator Documentation CD- ROM as
visaex6.cpp.
//****************************************************************************************

Table of Contents

Other manuals for Agilent Technologies E8257D

Questions and Answers:

Agilent Technologies E8257D Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8257D
CategoryInverter
LanguageEnglish

Summary

Getting Started with Remote Operation

Interfaces

Details various interface types (GPIB, LAN, RS-232, USB) for instrument communication.

IO Libraries and Programming Languages

Covers IO libraries and common programming languages for instrument control.

Using IO Interfaces

Using GPIB

Details GPIB interface for instrument connection and control, including setup and verification.

Using LAN

Explains LAN interface protocols (VXI-11, Sockets, Telnet, FTP) for remote instrument programming.

Programming Examples

Using the Programming Interface Examples

Introduces programming examples for remote control using various interfaces and languages.

GPIB Programming Interface Examples

Provides examples for GPIB programming and instrument control, including interface checks and queries.

LAN Programming Interface Examples

Covers LAN programming examples using VXI-11 and Sockets LAN protocols for instrument control.

Programming the Status Register System

Overview

Provides an overview of the status register system for monitoring signal generator conditions and errors.

Status Register Bit Values

Explains how each bit in a register is represented and how to enable/query bits using SCPI commands.

Accessing Status Register Information

Details methods for monitoring conditions, including polling and service request interrupts.

Status Byte Group

Describes the Status Byte Group, including the Status Byte Register and Service Request Enable Register.

Creating and Downloading Waveform Files

Creating and Downloading User-Data Files