EasyManuals Logo
Home>Agilent Technologies>Inverter>E8257D

Agilent Technologies E8257D User Manual

Agilent Technologies E8257D
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #188 background imageLoading...
Page #188 background image
178 Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide
Programming the Status Register System
Status Groups
Data Questionable Calibration Condition Register
The Data Questionable Calibration Condition Register continuously monitors the calibration status of
the signal generator. Condition registers are read only.
Data Questionable Calibration Transition Filters (negative and positive)
The Data Questionable Calibration Transition Filters specify which type of bit state changes in the
condition register set corresponding bits in the event register. Changes can be positive (0 to 1) or
negative (1 to 0).
Data Questionable Calibration Event Register
The Data Questionable Calibration Event Register latches transition events from the condition register
as specified by the transition filters. Event registers are destructive read- only. Reading data from an
event register clears the content of that register.
Table 4-11 Data Questionable Calibration Condition Register Bits
Bit Description
0
a
a.In the N5181A, E4428C, E8257D, and the E8663B, this bit is set to 0.
I/Q Calibration Failure. A 1 in this bit position indicates that the I/Q modulation calibration experienced a failure.
1
b
b.In the N5181A and N5182A, this bit is set to 0.
DCFM/DCΦM Zero Failure. A 1 in this bit position indicates that the DCFM/DCΦM zero calibration routine has
failed. This is a critical error. The output of the source has no validity until the condition of this bit is 0.
2−14 Unused. These bits are always set to 0.
15 Always 0.
Query: STATus:QUEStionable:CALibration:CONDition?
Response: The decimal sum of the bits set to 1.
Commands:
STATus:QUEStionable:CALibration:NTRansition <value> (negative transition), or
STATus:QUEStionable:CALibration:PTRansition <value> (positive transition), where <value> is
the sum of the decimal values of the bits you want to enable.
Queries:
STATus:QUEStionable:CALibration:NTRansition?
STATus:QUEStionable:CALibration:PTRansition?
Query: STATus:QUEStionable:CALibration[:EVENt]?

Table of Contents

Other manuals for Agilent Technologies E8257D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8257D and is the answer not in the manual?

Agilent Technologies E8257D Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8257D
CategoryInverter
LanguageEnglish

Related product manuals