Agilent InfinityLab LC Series Multicolumn Thermostats User Manual 108
7 Error Information
General Error Messages
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals.
A LOW signal input on pin 4 of the remote connector generates the error
message.
Probable cause Suggested actions
1 Leak detected in another module with a CAN
connection to the system.
Fix the leak in the external instrument before
restarting the module.
2 Leak detected in an external instrument with
a remote connection to the system.
Fix the leak in the external instrument before
restarting the module.
3 Shut-down in an external instrument with a
remote connection to the system.
Check external instruments for a shut-down con-
dition.
4 The degasser failed to generate sufficient
vacuum for solvent degassing.
Check the vacuum degasser for an error condi-
tion. Refer to the Service Manual for the degasser
or the pump that has the degasser built-in.