Agilent InfinityLab LC Series Multicolumn Thermostats User Manual 140
9 Maintenance
Replace the Module Firmware
Replace the Module Firmware
To upgrade/downgrade the module’s firmware carry out the following steps:
1 Download the required module firmware, the latest FW Update Tool and the
documentation from the Agilent web.
http://www.agilent.com/en-us/firmwareDownload?whid=69761
2 For loading the firmware into the module follow the instructions in the
documentation.
Module Specific Information
Module is a hosted module and always needs to be connected to a host with
matching firmware of same revision.
When
The installation of newer firmware might be necessary
• if a newer version solves problems of older versions or
• to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary
• to keep all systems on the same (validated) revision or
• if a new module with newer firmware is added to a system or
• if third party control software requires a special version.
Tools required Description
Agilent Lab Advisor software
Parts required # Description
1 Firmware, tools and documentation from Agilent web site
Preparations
Read update documentation provided with the Firmware Update Tool.