Agilent InfinityLab LC Series Multicolumn Thermostats User Manual 34
2 Site Requirements and Specifications
Performance Specifications
Performance Specifications
Performance Specifications G7116B
Table 3 Performance Specifications G7116B
Feature Specification Comment
Operating principle Thermostatted column compartment with dual, independent Pelt-
ier element.
Solvent pre-heating and still-air operation for reduction of chro-
matographic band broadening under UHPLC-conditions. Up to
four devices can be clustered and controlled by a single user inter-
face for additional flexibility.
1
Temperature range 20 °C below ambient (minimum 4 °C) to 110 °C
2
Temperature range
increment
Settable in steps of 0.1 °C
2
Temperature stabil-
ity
±0.03 °C
2
Temperature accu-
racy
±0.5 °C (with calibration at 40 °C)
2
Temperature preci-
sion
0.05 °C
2
Independent tem-
perature zones
2 (in single device)
Up to 8 in clustered configuration
1
Column capacity 8 columns of 100 mm length plus Quick Connect fittings or
pre-columns
4 columns of 300 mm length plus Quick Connect fittings or
pre-columns
Selection of columns by single optional integrated 8-column
selector valve (1300 bar)
Maximum of
32 columns of 100 mm length plus Quick Connect fittings or
pre-columns
16 columns of 300 mm length plus Quick Connect fittings or
pre-columns
with clustering
1
of four devices.