EasyManuals Logo
Home>Agilent Technologies>Thermostat>G7116B

Agilent Technologies G7116B User Manual

Agilent Technologies G7116B
203 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #115 background imageLoading...
Page #115 background image
Agilent InfinityLab LC Series Multicolumn Thermostats User Manual 115
7 Error Information
MCT Error Messages
WRONG VALVE
Error ID: 32130
Valve not supported in this module.
Probable cause Suggested actions
1 A valve head has been installed which is not
supported by G7116A.
Only valve heads up to 800 bar or with up to four
positions can be used in G7116A. Exchange the
valve head with a suitable one. For changing the
valve head follow the instructions Replace Valve
Heads.
NOT E
For G7116B:
G4232D (5067-4240: 2-position/10-port Valve 1300 bar)
G4231C (5067-4241: 2-position/6-port Valve 1300 bar)
G4234A (5067-4284: 6-position/14-port Valve 800 bar)
G4234C (5067-4273: 6-position/14-port Valve 1300 bar)
G4239C (5067-4233: 8-position/18-port Valve 1300 bar)
For G7116B Bio:
G5641A (5067-6682: 2-position/10-port Valve 1300 bar, bio)
For G7116A:
G4232C (5067-4283: 2-position/10-port Valve 800 bar)
G4231A (5067-4282: 2-position/6-port Valve 800 bar)
G4237A (5067-4279: 4-position/10-port Valve 800 bar)
For G7116A Bio-inert:
G5632A (5067-4132: 2-position/10-port Valve 600 bar, bio-inert)
G5631A (5067-4148: 2-position/6-port Valve 600 bar, bio-inert)
G5639A (5067-4134: 4-position/10-port Valve 600 bar, bio-inert)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7116B and is the answer not in the manual?

Agilent Technologies G7116B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7116B
CategoryThermostat
LanguageEnglish

Related product manuals