G2S A20 / G2S C20 User Manual V1.0
36 / 39
Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
user of the device should make sure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Conducted RF
IEC
61000-4-6
Radiated RF
IEC
61000-4-3
3 V
0.15 MHz
80 MHz
6 V in ISM
and amateur
radio bands
between
0.15 MHz
and 80 MHz
10 V/m
80 MHz to
2.7 GHz
3 V
0.15 MHz
80 MHz
6 V in ISM and
amateur radio
bands
between
0.15 MHz and
80 MHz
10 V/m
80 MHz to 2.7
GHz
Portable and mobile RF
communications equipment should
be used no closer to any part of the
device, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance
pd 17.1
pd 35.0
80 MHz to 800 MHz
pd 70.0
800 MHz to 2.5 GHz
Where
P
is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitter, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applied.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular /
cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the device is
used exceeds the applicable RF compliance level above, the device should be observed to
verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating the device.
b
Over the frequency range 150 kHz to 80 MHz, the field strengths should be less than 10
V/m.

Table of Contents

Other manuals for BMC G2S A20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BMC G2S A20 and is the answer not in the manual?

BMC G2S A20 Specifications

General IconGeneral
BrandBMC
ModelG2S A20
CategoryMedical Equipment
LanguageEnglish

Related product manuals