G2S A20 / G2S C20 User Manual V1.0
37 / 39
Recommended separation distances between portable and mobile RF
communications equipment and the device
The device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the device as recommended
below, according to the maximum output power of the communications equipment.
Rated
maximum
output of
transmitter
W
150 kHz 80 MHz
pd 17.1
80 MHz 800 MHz
pd 35.0
800 MHz 2.5 GHz
pd 70.0
0.01
0.12
0.04
0.07
0.1
0.37
0.12
0.23
1
1.17
0.35
0.70
10
3.70
1.11
2.22
100
11.7
3.50
7.00
Note 1: At 80 MHz and 800 MHz, the higher frequency range applied.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.

Table of Contents

Other manuals for BMC G2S A20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BMC G2S A20 and is the answer not in the manual?

BMC G2S A20 Specifications

General IconGeneral
BrandBMC
ModelG2S A20
CategoryMedical Equipment
LanguageEnglish

Related product manuals