Property Description
Alarm text
Disconnection
Priority High
Alarm condition A Disconnection alarm will be given when the measured
flow exceeds the expected leakage flow at the set Pressure
for 15 seconds.
Possible cause
• Too high leakage in the patient circuit.
• The patient has removed the mask.
• Circuit disconnection.
Reset criteria
The leakage is back within limits.
Ventilator action The Vivo 2 will continue treatment according to the current
settings
Setting range
• On
• Off
7.3.10 High EPAP Alarm
Property Description
Alarm text
High EPAP
Priority
Medium
Alarm condition A High EPAP alarm will be given when EPAP has gone
30% above the set value for 3 breaths.
Possible cause
• Blocked leakage port.
• Too short expiratory time.
• Changes in airway resistance and or compliance.
Reset criteria
EPAP has gone below the alarm limit (lower than 30%
above the set value).
Ventilator action The Vivo 2 will continue treatment according to the current
settings.
Setting range
• On
• Off
100 Alarms
User manual
Doc. 007229 D-3