Property Description
Alarm text
Disconnection
Priority High
Alarm condition A Disconnection alarm will be given when the measured
flow exceeds the expected leakage flow at the set Pressure
for 15 seconds.
Possible cause
Too high leakage in the patient circuit.
The patient has removed the mask.
Circuit disconnection.
Reset criteria
The leakage is back within limits.
Ventilator action The Vivo 2 will continue treatment according to the current
settings
Setting range
On
Off
7.3.10 High EPAP Alarm
Property Description
Alarm text
High EPAP
Priority
Medium
Alarm condition A High EPAP alarm will be given when EPAP has gone
30% above the set value for 3 breaths.
Possible cause
Blocked leakage port.
Too short expiratory time.
Changes in airway resistance and or compliance.
Reset criteria
EPAP has gone below the alarm limit (lower than 30%
above the set value).
Ventilator action The Vivo 2 will continue treatment according to the current
settings.
Setting range
On
Off
100 Alarms
User manual
Doc. 007229 D-3

Table of Contents

Other manuals for Breas Vivo 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 3 and is the answer not in the manual?

Breas Vivo 3 Specifications

General IconGeneral
BrandBreas
ModelVivo 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals