7.3.6 Low Minute Volume Alarm
Property Description
Alarm text
Low MV
Priority High
Alarm condition A Low Minute Volume alarm will be given when the
minute volume is below the set alarm limit for
15 1 seconds.
Possible cause
• Disconnection of patient circuit.
• Mismatch between pressure setting and alarm setting.
• Leakage from the mask or other components of the patient
circuit.
• Decreased breath rate.
Reset criteria The minute volume goes above the alarm limit.
Ventilator action The Vivo 2 will continue treatment according to the cur-
rent settings.
Setting range
• 1 l. to 30 l.
• Off
Setting resolution 0.5 l.
Alarms 97
Doc. 007229 D-3 User manual