10 Accessories
The accessories described in this section, together with the medical electric equipment
defined in chapter 3.1 Main Components, page 30, constitute the Vivo 2 medical electric
system.
CAUTION!
Responsibility for System
Accessory equipment connected to the analogue and digital interfaces must be certi-
fied according to the respective IEC standards (e.g. IEC 60950 or IEC 62368 for
data processing equipment and IEC 60601-1 for medical equipment). Furthermore
all configurations must comply with the valid version of the system standard IEC
60601-1. Anybody who connects additional equipment to the signal input part or sig-
nal output part is configuring a medical system, and is therefore responsible for
ensuring that the system complies with the requirements of the valid version of the
system standard IEC 60601-1. If in doubt, consult the technical service department
or your local representative.
A PC, when connected to the ventilator, shall comply with IEC 62368-1, IEC
60950-1 or IEC 60601-1.
10.1 Accessories/Consumables
Circuit: Patient circuit with leakage
Function: Delivers air to the patient,
applied part
Part No: 005060
Accessories 121
Doc. 007229 D-3 User manual