Supplemental Oxygen Supply Requirements
Property Requirement
Maximum flow
30 l/min.
Maximum pressure 100 kPa
Supply source Source equipped with rotameter.
Examples of supply sources:
• Oxygen cylinder
• Central oxygen supply
• Oxygen concentrator
Connector The oxygen source shall be equipped
with the Breas Low Pressure Oxygen
Adapter, art. no. 005032
Connect the Oxygen supply
1 Connect the oxygen adapter (article no. 005032) to the oxygen supply’s tube.
2 Connect the oxygen adapter to the oxygen port at the back of the Vivo 2. See 3.3
Ventilator Back , page 33 for detailed information.
3 If using a device for monitoring the oxygen saturation, connect it according to the
supplier instructions.
CAUTION!
Do not add Oxygen before therapy has been started.
6.6.5 Using the Remote Alarm Unit
Information about safety, warnings, product description, installation, usage, cleaning,
maintenance and technical specifications can be found in the user instruction for
Remote Alarm.
The Remote Alarm enables care providers and clinical personnel to monitor the Vivo 2
alarms remotely. The Remote Alarm forwards alarms from the Vivo 2.
1 Connect the Accessory box to the Communication port at the back of the Vivo 2.
80 How to Use the Vivo 2
User manual
Doc. 007229 D-3