8400-0181-OM Rev C 185 and 185B Installation & Technical
26
Continuous Output
If the Continuous Output for the serial port was selected COnt=yes, then the 185 will transmit
weight repeatedly. The format is determined by the setting of the tyPE= parameter:
0 = SM
1 = SB-400
SMA Format
<lf><s><r><n><m><f><xxxxxx.xxx><uuu><cr>
where:
lf = Line Feed (hex 0A)
s = Flag Z for center of zero
O for overcapacity
U for negative overcapacity
e for weight not displayed, or space
r = Range 1 (single range)
n = Mode G for gross, N for net
m = Motion M for motion, space for stable
f = Custom Fla
lwa
s a space
xxxxxx.xxx = Weight ten digits, including decimal point, right justified
uuu = Units lb^, kg^, oz^, or g^^, ^ = space
cr = Carriage Return (hex 0D)
SB-400 Format
<s><xxxxxx><d><_><uu><_><m><_><cc><_><cr>
where:
s = Sign "-" = negative, “ “ (space) for zero and above
xxxxxx.xxx = Weight Six digits
d = Decimal point Embedded in the weight, if enabled in setup
_ Space
uu = Units LB, KG, OZ, or G^, ^ = space
m = Mode G for gross or N for net
cc = Weight Status
OC for overcapacity
MO for unstable
CZ for center of zero
EE for weight not displayed, or two spaces
cr = Carria
e Return
hex 0D