Instructions for use Isolette 8000 plus SW 5.n 49
Assembly and preparation
WARNING
Risk of death or serious injury
Latches can come loose leading to disconnecting of devices from the rail.
To avoid injury or damage to the device, use only Dräger-approved latches,
components, and accessories with the Fairfield-compatible rail system.
It is also important to secure the latch properly to the rail as identified in the
manual.
Periodically check the operation of the latch to hold securely the accessory
being used. When the latch is properly locked, both visual and physical
feedback are provided. The toggle handle snaps into an 8 o’clock position on the
latch body, when locked.
Total rail system weight should not exceed 13.6 kg (30 lb); 6.8 kg (15 lb) per
side.
For weight limitations for accessories and components for the Isolette 8000 plus,
see Technical Data on page 167.
6.4 Installing the weighing system (option)
6.4.1 Standard scale assembly
1 Rotate the locking knobs, and open the front access panel of the incubator.
2 Remove the mattress from the incubator.
3 Orient the scale so that the scale cable is on the right-hand side of the incubator
towards the sensor module assembly.
4 Place the scale (A) in the incubator on the bed (B).

Table of Contents

Other manuals for Dräger Isolette 8000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette 8000 plus and is the answer not in the manual?

Dräger Isolette 8000 plus Specifications

General IconGeneral
BrandDräger
ModelIsolette 8000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals