Instructions for use Isolette 8000 plus SW 5.n 89
Operation
8.2.3 Setting the humidity set point in manual mode
If no keys are pressed within 15 seconds of making any selection, the screen
automatically reverts to the previous display.
To set the humidity set point:
1 Activate the humidification system in manual mode (see Selecting the humidity
mode on page 87).
2 Adjust the humidity set point.
To raise the humidity set point within a range of 30% to 95% in 1%
increments, press and hold the Up Arrow key (A).
To lower the humidity set point within a range of 95% to 30% in 1%
decrements, press and hold the Down Arrow key (B).
3 To confirm the humidity setting, press the Home button (C).
Device returns to display 1.
4 Press the Keypad Lock key (D).
✓ Keypad is locked.
8.3 Setting the servo control oxygen (option)
In oxygen control mode, the oxygen sensors and control valve module control the
oxygen concentration (from 21% to 65%). The alarm limit default is ± 3% from the
set point. If the oxygen concentration rises above or falls below the ± 3% limit,
acoustic and optical alarm signals are issued. Also, the message Low Oxygen %
or High Oxygen % is displayed in the Trend/Alarm window.
WARNING
Risk of death or serious injury
The oxygen concentration guide is valid only for manual control oxygen usage.
► When using the servo control oxygen system during oxygen administration, do
not use the oxygen concentration guide on the rear of the device.
If arterial oxygen levels cannot be maintained when the oxygen control setting is
set to maximum, the attending physician should prescribe alternate means of
oxygenation.
055