Instructions for use Isolette 8000 plus SW 5.n 61
Getting started
7 Getting started
7.1 System start-up
WARNING
Risk due to incorrect mains voltage or missing protective grounding
If the device is connected to a power socket with incorrect mains voltage or a power
socket without a protective grounding, an electric shock may occur.
► Connect the device only to power sockets with correct mains voltage and a
protective grounding.
WARNING
Risk of reduced level of safety
Connecting other devices to auxiliary power sockets on the I8000 plus may lead to
a reduced level of safety.
► Comply with the requirements of IEC 60601-1 when connecting other devices to
auxiliary power sockets on the Isolette 8000 plus.
► Do not use auxiliary power sockets that are separate from the Isolette 8000 plus
WARNING
Risk of patient injury or device failure
If the device loses its power source, the patient could be injured or the device
damaged.
► The incubator must be connected to an appropriate power source when loss of
power source would result in an unacceptable risk.
Prerequisites:
– The device has been reprocessed and assembled ready for operation.
Procedure:
Always perform the following test steps before operating the device.
1 Verify that the power cable from the trolley is firmly attached to the receptacle
(A) under the incubator shell assembly.
020