Technical data13
13.6 FLIR A310
P/N: 48201-1101
Rev.: 22369
General description
The FLIR A310 offers an affordable and accurate temperature measurement solution for anyone who
needs to solve problems that need built in “smartness” such as analysis, alarm functionality and autono-
mous communication using standard protocols. The FLIR A310 also has all the necessary features and
functions to build distributed single- or multi-camera solutions utilizing standard Ethernet hardware and
software protocols.
The FLIR A310 also has built in support to connect to industrial control equipment such as PLCs, and
allows for sharing of analysis and alarm results and simple control using the Ethernet/IP and Modbus
TCP field bus protocols.
Key features:
• Support for EthernetIP field bus protocol (analyze, alarm, and simple camera control).
• Support for Modbus TCP field bus protocol (analyze, alarm, and simple camera control).
• Built-in extensive analysis functionality.
• Extensive alarm functionality, as a function of analysis and more.
• On schedule: file sending (FTP) or e-mail (SMTP) of analysis results or images.
• On alarms: file sending (FTP) or e-mail (SMTP) of analysis results or images.
• MPEG-4 streaming.
• PoE (Power over Ethernet).
• Built-in web server.
• General purpose I/O.
• 100 Mbps Ethernet (100 m cable, wireless, fiber, etc.).
• Synchronization through SNTP.
• Composite video output.
• Multi-camera utility software: FLIR IP Config and FLIR IR Monitor included.
• Open and well-described TCP/IP protocol for control and set-up.
• 16-bit 320 × 240 pixel images at 7–8 Hz, radiometric.
• Lenses: 25° included, 15° and 45° optional.
Typical applications:
• Safety with temperature alarms (multi-camera applications), fire prevention, critical vessel monitor-
ing, and power utility asset management.
• Volume-oriented industrial control (multi-camera installation is possible).
Imaging and optical data
IR resolution 320 × 240 pixels
Thermal sensitivity/NETD < 0.05°C @ +30°C (+86°F) / 50 mK
Field of view (FOV)
25° × 18.8°
Minimum focus distance 0.4 m (1.31 ft.)
Focal length 18 mm (0.7 in.)
Spatial resolution (IFOV)
1.36 mrad
Lens identification Automatic
F-number 1.3
Image frequency 30 Hz
Focus Automatic or manual (built in motor)
Zoom 1–8× continuous, digital, interpolating zooming on
images
Detector data
Detector type Focal Plane Array (FPA), uncooled
microbolometer
Spectral range
7.5–13 µm
Detector pitch 25 µm
Detector time constant Typical 12 ms
#T559498; r.22370/22370; en-US
37