Preventive maintenance: Battery maintenance
5-14 Dash 3000/4000/5000 2000966-542D
5. Select Wake Up Lithium Ion Battery in SLOT A. Then, the
patient monitor applies a “wake-up” shot of current for up to 210
seconds.
6. Watch for the message, Battery awake, run main code to charge.
The battery should now be “awake.”
If the “Battery awake, run main code to charge” message
appears but the battery is NOT awake, the battery may be in a
deep discharge state. You should repeat step 5 at least four more
times to give the battery additional “wake-up” shots of current.
7. Quickly reboot (or power cycle) the patient monitor to begin charging
the batteries. The patient monitor’s illuminated yellow-colored
CHARGING STATUS LED indicates that the battery is being
charged.
NOTE
The patient monitor will not charge the battery while it is
running the Boot Loader program.
How to replace the batteries
WARNING
EXPLOSION OR FIRE — Using non-recommended
batteries could result in injury/burns to the patient or
users.
Only use batteries recommended or manufactured by GE.
The warranty can be voided if non-recommended
batteries are used.
1. Open the battery doors on the left side of the patient monitor, along
the bottom.
2. Remove the faulty batteries.

Table of Contents

Other manuals for GE Dash 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 3000 and is the answer not in the manual?

GE Dash 3000 Specifications

General IconGeneral
BrandGE
ModelDash 3000
CategoryMonitor
LanguageEnglish

Related product manuals