Configuration: Service Mode settings
2000966-542D Dash 3000/4000/5000 4-11
Service Mode settings
From the Main Menu,
1. Select MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the control to select the day and month from
patient monitor screen with leading zeros. (e.g. July 4 = 0407)
Set Unit Name
Up to seven characters are used to identify the care unit. These
characters display at the top right of the screen immediately preceding
the bed number.
1. Enter Service Mode (See “Service Mode settings” on page 4-11) and
select MONITOR SETTINGS > UNIT NAME.
2. Use the Trim Knob control to select and change each character. Up
to seven characters may be entered.
3. Select SET UNIT NAME and press the Trim Knob control to exit.
Set Bed Number
The bed number identifies a particular patient bed. Up to five characters
are used to identify bed number. This number displays at the top right of
the screen.
1. Enter service mode (See “Service Mode settings” on page 4-11) and
select MONITOR SETTINGS > UNIT NAME > SET BED
NUMBER.
2. Use the Trim Knob control to select and change each character. Up
to five characters may be entered.
3. Select SET BED NUMBER and press the Trim Knob control to exit.
↓ ↑
040A
↓ ↑
041A

Table of Contents

Other manuals for GE Dash 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 3000 and is the answer not in the manual?

GE Dash 3000 Specifications

General IconGeneral
BrandGE
ModelDash 3000
CategoryMonitor
LanguageEnglish

Related product manuals