Configuration: Boot Code settings
4-22 Dash 3000/4000/5000 2000966-542D
1. Activate Boot Code (“Boot Code settings” on page 4-18) and select
SET CONFIGURATION menu option.
2. Select the IntelliRate menu option.
3. Select enabled to activate this option, or disabled to deactivate this
option.
4. Select Exit and wait for the patient monitor to display the message,
Writing settings to Flash config mem...
5. Manually reboot the patient monitor to implement this change.
6. Refer to the operator’s manual to configure the patient monitor to
use this feature.
Analog Out Buzz
Select Analog Out Buzz to identify an ECG or an invasive BPleads off
condition via a supported analog out device. The default is 0 volts.
1. Activate Boot Code (“Boot Code settings” on page 4-18) and select
SET CONFIGURATION menu option.
2. Select the Analog Out Buzz menu option.
3. Select 60Hz sine wave to display a 60 Hz sine wave, or select 0 volts
to display a flat line.
4. Select Exit and wait for the patient monitor to display the message,
Writing settings to Flash config mem...
5. Manually reboot the patient monitor to implement this change.
Completion
The patient monitor is now ready for normal operation. At this time,
perform procedures in “Functional and electrical safety checks” on
page 8-1.

Table of Contents

Other manuals for GE Dash 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 3000 and is the answer not in the manual?

GE Dash 3000 Specifications

General IconGeneral
BrandGE
ModelDash 3000
CategoryMonitor
LanguageEnglish

Related product manuals