Configuration: Boot Code settings
2000966-542D Dash 3000/4000/5000 4-19
Set CIC and QS protocol
The CIC and QS Protocol default setting is Seg50/51 (Segment 50/51)
and should not be changed. However, future products may require
Hilltop protocol. Use the Boot Code Service Menu to change the CIC
Protocol and QS Protocol.
Set MUSE system protocol
The patient monitor transmits 12SL and ACI-TIPI data over the optional
CARESCAPE Network to the MUSE System and ST Guard. The formats
used for this process are Hilltop and Segment 50/51. Since the Segment
50/51 format does not support 500 Hz ECG data or ACI-TIPI, records
originally stored on the MUSE System in Hilltop format cannot display
at the ST Guard. The default setting is Hilltop.
1. Activate Boot Code (“Boot Code settings” on page 4-18) and select
SET CONFIGURATION menu option.
2. Select MUSE System Protocol then choose Hilltop or Seg50/51.
Hilltop—Sends 12SL records to the MUSE System in Hilltop
format. The 12SL records contain 500 samples per second of ECG
data. ST Guard will not be able to retrieve 12SL records stored on
the MUSE System in Hilltop format.
Seg50/51—Sends 12SL records to the MUSE System in Segment
50/51 format with MAC Rhythm statements. The 12SL records
stored on the MUSE System contain 240 samples per second of
ECG data. ST Guard can retrieve 12SL records from the MUSE
System. ACI-TIPI data can not be retrieved at the MUSE System
if this protocol is selected.
3. Select Exit and wait for the patient monitor to display the message,
“Writing settings to Flash config mem...”
4. Manually reboot the patient monitor to implement this change.
NOTE
Before you can transmit data to the MUSE System and ST
Guard, you must set up the MUSE system’s Location ID and Site
Number.
Do not configure MUSE system’s location at Site 0. Site 0
appears invalid to MUSE and any message coming from a device
on Site 0 is dumped.
5. Refer to “Service Mode settings” on page 4-11 and select MONITOR
SETTINGS > SET LOCATION, then select SET SITE NUMBER.
Transcutaneous Pace Blank Length