Field replaceable units: Replace main unit parts
2000966-542D Dash 3000/4000/5000 7-43
5. Remove the four screws anchoring the speaker to the frame. Remove
the speaker from the unit.
6. Install the new speaker in the unit.
7. Fasten the speaker with the four screws.
8. Connect the speaker cable connector to the processor board.
9. Reassemble the patient monitor in reverse order.
NOTE
When securing the thumb screw on the writer flex, first finger-
tighten the thumb screw. Then, use a flat-head screw driver to
tighten the screw an additional 1/4 turn.
10. Complete the procedures in “Recommended checkout” on page 7-52.